Calibration EMF and Pressure Sensor

Calibration EMF and Pressure Sensor

This calibration procedure applies to instruments of the laboratory for physical geography of Utrecht University. Be aware that the so called afdeelfactor does not apply to all pressure sensor calibrations. Often, the afdeelfactor is already taken into account in the calibration coefficients provided by the lab and should therefore be left out of the calibration below.

Calibration EMF 147

u = \frac{U_{\mathrm{cr10}}}{\mathrm{overdracht}} -\frac{\mathrm{offset}}{\mathrm{overdracht}}

Example

X-direction
Afdeelfactor 0.2482
Overdracht 255.53
Offset -5.23

u_{x} = 0.0039 \times U_{\mathrm{cr10}} + 0.020

Y-direction
Afdeelfactor 0.2485
Overdracht 254.56
Offset 7.01

u_{y} = 0.0039 \times U_{\mathrm{cr10}} - 0.028

Calibration Pressure Sensor

p_{\mathrm{water}} = \frac{\frac{U_{\mathrm{cr10}}}{\mathrm{afdeelfactor}} - \mathrm{Constante}}{\mathrm{Xcoefficient}} - p_{\mathrm{air}}
Use the Usensor = f(p) regression!

Example

pressure sensor 2774
Constant -6378.39 mV
Xcoefficient 6.653486 mV/mbar
Afdeelfactor 0.2479

p_{\mathrm{water}} = 0.6063 \times U_{\mathrm{cr10}} + 958.65 - p_{\mathrm{air}}

where the unit of water pressure is millibar (mbar)

1 m fresh water:

\rho_{\mathrm{water}}gh = 1000 \times 9.81 \times 1.0 = 9810 \mathrm{\,N/m^2} = 98.1 \mathrm{\,\, mbar}

so: 1 mbar = 0.0102 m of fresh water

m salt water:

\rho_{\mathrm{water}}gh = 1025 \times 9.81 \times 1.0 = 10055 \mathrm{\,N/m^2} = 100.55 \mathrm{\,\, mbar}

so: 1 mbar = 0.0099 m water

Using this, the calibration becomes approximately:

\eta_{\mathrm{water}} = 0.006063U_{\mathrm{cr10}} + 9.5865 - 0.01 p_{\mathrm{air}}

where the unit of the water height is meter (m), and the unit of air pressure is millibar (mbar)

Powered by Drupal - Modified by Danger4k